Testosterone, Total
Tube Type
| Test ID |
|
LAB124 |
EPIC Order
|
|
TESTOSTERONE |
| CPT Code(s) |
|
84403 |
| Group/Individual Test |
|
Individual |
| Laboratory |
|
Core Laboratory |
| Tube Station |
|
30 |
| Specimen |
Routine: |
4 mL blood/gold SST tube/red plain tube also acceptable |
| |
Micro: |
0.9 mL blood/gold or red top bullet |
| Stability |
|
Unspun: 24 hours
Spun/separated, Room Temp: 48 hours
Spun/separated, Refrigerator: 7 days
Spun/separated, Frozen: 6 months |
| Availability |
|
24 hrs/day |
| Turnaround Time |
Routine: |
2 hrs |
|
Stat: |
90 mins |
| Reference Range |
Normal: |
see table below |
|
|
Reference Ranges: Testosterone ng/dL (effective 7/13/2021)
| Patient Age |
Female |
Male |
| 0-2 years |
<34 |
<263 |
| 2-10 years |
<12 |
<11 |
| 11-12 years |
<28 |
<479 |
| 12-13 years |
<28 |
<488 |
| 13-14 years |
<28 |
8-550 |
| 14-15 years |
<28 |
9-535 |
| 15-16 years |
<28 |
66-757 |
| 16-21 years |
12-43 |
228-710 |
| 22-50 years |
8-35 |
197-670 |
| +50 years |
<36 |
188-684 |
|
| |
- Testing is performed using Siemens Atellica TST II Assay. Different test methods should not be used interchangeably. Heterophilic antibodies may cause interference. High doses of biotin may cause test interference.
- Cross reactivity can occur with Nandrolone decanoate, 11 beta-hydroxytestosterone and 11-ketotestosterone. Steroid based medications could potentially cause falsely elevated values.
|
Reviewed by J. Carr on January 02, 2026