Estrogen, Serum (Estradiol)

Tube Type = Gold Top

Tube Type

Test ID LAB523
EPIC Order ESTRADIOL (ESTROGEN) LEVEL
CPT Code(s) 82670
Group/Individual Test Individual
Laboratory Core Laboratory
Tube Station 30
Specimen Routine: 4 mL blood/Gold top serum separator tube / Also acceptable serum Red Top Tube
Micro: 500µL blood/Yellow top serum separator bullet/red top
Stability Spun/Separated: Room Temperature: 20 hours
Refrigerated: 48 hours
Frozen: 6 months
Availability Routine: 24 hrs/day
Turnaround Time Routine: 2 hours
Reference Range Normal: see Table below
Comments
Reference Ranges: Serum Estrogen*
(pg/ml)
Male: <39.8
Female (Postmenopausal): <32.2
Female (Ovulating): (pg/ml)
follicular phase 19.5-144.2
periovulatory (±3 days) 63.9-356.7
luteal phase 55.8-214.2
*Reference Ranges effective 7/13/2021

Testing is performed using Siemens Atellica Estradiol (eE2) Assay.  Different test methods should not be used interchangeably.  Heterophilic antibodies may cause interference. 

The drug Fulrestrant may cause falsely elevated estradiol results in immunoassays.  There is potential for cross-reactivity with steroid-based medications with chemical structures like estradiol.

Results of this assay should be interpreted in conjunction with the patient’s history, clinical presentation, and other findings.  Additional testing by alternate methods (LC-MS for example) maybe indicated to confirm results.